Dimension CTNI-In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and... — FDA record Z-1341-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 18, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Dimension CTNI-In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444905/RF421C
Reason
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
113 units; Extended: 52 units
Agency-reported status
Terminated
Source record identifier
Z-1341-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00842768013836GA232320221119 Lot Number: GA2323 Exp. Date: 11-19-2022 Recall Extended 9/21/22: UDI: 00842768013836BA308820230329 Lot Number: BA3088 Exp. Date: 2023-03-29

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, IL MN, MS, NC, ND, NH, OK, SD, TN, TX, WA and the countries of Argentina, Brazil, Canada, France, Germany, India, Mexico, Poland, Portugal, Slovakia, Spain.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 13, 2022
Recall initiation date
May 18, 2022
Center classification date
Jul 6, 2022
Termination date
Feb 13, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.