MINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 2.75" PG Catalog Number: 45-756 — FDA record Z-1339-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 30, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 2.75" PG Catalog Number: 45-756
Reason
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
Classification
Class II
Recalling firm
Angiodynamics, Inc.
Quantity
5176 units
Agency-reported status
Ongoing
Source record identifier
Z-1339-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 15051684022996 UPN: H965457561 Lot Number: 5796084 5796085 5796086 5796087 5796088 5796089 5796090 5796659 5796660 5796661 5796662 5796663 5796666 5797456 5797465 5798836 5798841 5798842 5799620 5800404 5803526 5804396

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 27, 2024
Recall initiation date
Jan 30, 2024
Center classification date
Mar 21, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records