HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient'... — FDA record Z-1329-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Mar 11, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery
- Reason
- Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.
- Classification
- Class II
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 80 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1329-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | 123510000293, 123510000294, 123510000295, 123510000296, 123510000298, 123510000299, 123510000300, 123510000764, 123510000765, 123510000766, 123510000767, 123510000768, 123510000769, 123510000770, 123510000771, 123510000772, 123510000773, 123510000800, 123510000801, 123510000802, 123510000803, 123510000804, 123510000805, 123510000806, 123510000807, 123510000808, 123510000809, 123510000297, 123350001033, 123350001034, 123350001035, 123350001036, 123350001040, 123350001041, 123350001042, 123360000294, 123360000295, 123360000297, 123360000298, 123360000299, 123360000300, 123360000301, 123360000302, 123360000303, 123360000304, 123360000733, 123360000734, 123360000735, 123360000736, 123360000737, 123360000738, 123360000739, 123360000740, 123360001972, 123360001973, 123360001974, 123360001975, 123360001976, 123360001977, 123360001978, 123360001979, 123360001980, 123360001981, 123350000968, 123350000969, 123350000970, 123350000971, 123350000972, 123350000973, 123350000974, 123350000975, 123350000987, 123350000988, 123350000989, 123350000990, 123350000991, 123350000992, 123350000993, 123350000994, 123350001032 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the state of Florida.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Mar 27, 2024
- Recall initiation date
- Mar 11, 2024
- Center classification date
- Mar 20, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.