Recall archive / FDA device enforcement
Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Syste... — FDA record Z-1325-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 22, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems 10135936RH, 08255790RH
- Reason
- These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
- Classification
- Class I
- Recalling firm
- Medline Industries, LP
- Quantity
- 511
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1325-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | 10135910RH UDI-DI 10197344026500 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905; 10135936RH UDI-DI 10197344026517 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Distribution US nationwide.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Feb 18, 2026
- Recall initiation date
- Dec 22, 2025
- Center classification date
- Feb 6, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.