Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense Webster SOUNDSTAR eco Diagnostic... — FDA record Z-1324-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 22, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on GE Systems 10439072RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on Siemens Systems 10438577RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter, Use on Siemens Systems 10439011RH
Reason
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Classification
Class I
Recalling firm
Medline Industries, LP
Quantity
94
Agency-reported status
Ongoing
Source record identifier
Z-1324-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability10439072RH UDI-DI 10197344157297 Lots EP250724 EP250808; 10439236RH UDI-DI 10197344157303 Lots EP250724 EP250808; 10438577RH UDI-DI 10197344157273 Lots EP250724 EP250808; 10439011RH UDI-DI (case) 20197344157287¿ (ea) 10197344157280¿ Lots EP250724 EP250808

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Distribution US nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 18, 2026
Recall initiation date
Dec 22, 2025
Center classification date
Feb 6, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.