Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967 — FDA record Z-1321-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 9, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967
Reason
Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.
Classification
Class II
Recalling firm
Preat Corp
Quantity
561 units
Agency-reported status
Ongoing
Source record identifier
Z-1321-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel Number: 9000967 UDI-DI code: 00842092165867 Lot Numbers: 235650, 234294, 233667, 233885, 233439, and 231489

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S. Nationwide distribution in the states of AZ, OH and PA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 27, 2024
Recall initiation date
Feb 9, 2024
Center classification date
Mar 20, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.