Abbott Point of Care Control Control i-STAT¿ Level 3 1.7 mL — FDA record Z-1315-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 9, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Abbott Point of Care Control Control i-STAT¿ Level 3 1.7 mL
Reason
Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
Classification
Class II
Recalling firm
Mckesson Medical-Surgical Inc. Corporate Office
Quantity
2 units
Agency-reported status
Ongoing
Source record identifier
Z-1315-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalogue Number: 06F1401; UDI/DI: 54749000661; Lot No: 121164;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of AZ, PA and TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 27, 2024
Recall initiation date
Jan 9, 2024
Center classification date
Mar 19, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.