Recall archive / FDA device enforcement
Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50, — FDA record Z-1310-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Feb 24, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50,
- Reason
- Certain Iradimed 1057 Syringe Adapter Sets, used with the MRidium MRI infusion pump, have a problem relating to a defective injection molded part which can limit the effectiveness of the syringe venting function which then can result in occlusion.
- Classification
- Class II
- Recalling firm
- Iradimed Corporation
- Quantity
- 3285 boxes of 50 devices
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1310-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI di 10856038001048, Lot Numbers: 21K50N, exp. 11/30/2023; 22A03N, exp. 01/31/2024; 22A59N, exp. 2/29/2024; 22B54N, exp. 2/29/2024; 22D15N, exp. 4/30/2024; 22D28N, exp. 4/30/2024; 22D58N, exp. 4/30/2024; 22I17N, exp. 9/30/2024; 031851, exp. 9/30/2024; 22J46N, exp. 10/31/2024; 032198, exp. 11/30/2024; 22L02N,exp. 12/31/2024; 032822, exp. 1/31/2025 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Canada, Chile, Denmark, Germany, Great Britain, Hong Kong, Indonesia, Italy, Korea, Kuwait, Malaysia, Netherlands, Nicaragua, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Singapore, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Apr 5, 2023
- Recall initiation date
- Feb 24, 2023
- Center classification date
- Mar 30, 2023