Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments. — FDA record Z-1307-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 13, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments.
Reason
Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response; adverse local tissue reactions may occur related to cytotoxicity with the nonconforming rod and surrounding tissue.
Classification
Class II
Recalling firm
K2M, Inc.
Quantity
53 devices (19 US; 34 OUS)
Agency-reported status
Ongoing
Source record identifier
Z-1307-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 10888857326804 Lot PCMW

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution to FL GA NE NV NY OH OK SC Foreign distribution to Canada, EU, and Hong Kong.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 6, 2022
Recall initiation date
Jun 13, 2022
Center classification date
Jun 30, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records