natus neurology, Sleep Supplies Start Kit with Single Use Supplies for Embla EX Series Amplifiers, REF 024252 — FDA record Z-1306-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 7, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
natus neurology, Sleep Supplies Start Kit with Single Use Supplies for Embla EX Series Amplifiers, REF 024252
Reason
Natus has become aware that its third-party supplier Parker Laboratories Inc. is conducting a voluntary recall related to their Tensive Conductive Adhesive Gel, Ref. 22-60 Lot Number A0623009. They have received five complaints. These customers reported that the gel is too thick, is discolored (darker than usual), and in one instance, patients reported irritation after procedures using the gel.
Classification
Class II
Recalling firm
Natus Neurology Inc
Quantity
35 units
Agency-reported status
Ongoing
Source record identifier
Z-1306-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI x, Lot Numbers: 614922, 614923, 614921, 122923, 628166

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 27, 2024
Recall initiation date
Feb 7, 2024
Center classification date
Mar 19, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.