Product Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applicable Pr... — FDA record Z-1303-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 11, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Product Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Calibrators: The pTau 217 plasma concentration of a specimen is automatically calculated from the calibration curve, which is also automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G pTau 217 Plasma Calibrators: Liquid (Frozen), 1x1.5 mL (5 Concentrations) CAL 1 0 pg/mL pTau 217 Plasma Calibrator CAL 2 0.250 pg/mL pTau 217 Plasma Calibrator CAL 3 1.000 pg/mL pTau 217 Plasma Calibrator CAL 4 5.000 pg/mL pTau 217 Plasma Calibrator CAL 5 10.000 pg/mL pTau 217 Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 300, 0.05% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Reason
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Classification
Class II
Recalling firm
Fujirebio Diagnostics, Inc.
Quantity
117 units
Agency-reported status
Ongoing
Source record identifier
Z-1303-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel/Catalog Number: 81557; UDI-DI: 05414470815579; Lot Numbers: D5C6094U, D5C6125U;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 11, 2026
Recall initiation date
Dec 11, 2025
Center classification date
Feb 5, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.