RayStation 11B — FDA record Z-1299-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 10, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
RayStation 11B
Reason
This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
Classification
Class II
Recalling firm
RAYSEARCH LABORATORIES AB
Quantity
26
Agency-reported status
Terminated
Source record identifier
Z-1299-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI 0735000201042620211208 Software version numbers 12.0.0.932, 12.1.0.1221, 12.0.3.68

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of NJ and TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 6, 2022
Recall initiation date
Jun 10, 2022
Center classification date
Jun 28, 2022
Termination date
Jun 11, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.