CARTO VIZIGO Bi-Directional Guiding Sheath, REF D128502 The CARTO VIZIGOTM Bi-Directional Guiding Sheath is designed ... — FDA record Z-1290-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 16, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
CARTO VIZIGO Bi-Directional Guiding Sheath, REF D128502 The CARTO VIZIGOTM Bi-Directional Guiding Sheath is designed to provide accessibility and maneuverability in the cardiac anatomy. The steerable sheath is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A side port with a three-way stopcock is provided for air or blood aspiration, and fluid infusion. The handle is equipped with a rotating collar to deflect the tip clockwise = 180¿ and counterclockwise = 180¿. The steerable sheath features distal vent holes to facilitate aspiration and minimize cavitation, and a radiopaque tip marker to allow fluoroscopic visualization. The sheath has electrodes on the outer surface to allow the sheath to interface with compatible CARTO" 3 Systems.
Reason
Due to a manufacturing issue, device under process validation phase were inadvertently mixed into the main manufacturing process.
Classification
Class II
Recalling firm
Biosense Webster, Inc.
Quantity
853 units
Agency-reported status
Ongoing
Source record identifier
Z-1290-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct No. D-1385-02 UDI-DI code: 10846835016277 Lot Numbers: 60000282 60000282

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S. Nationwide distribution including in the states of FL, IL, AL, NY, PA, SC, AZ, MI KY, NJ, WI, IN, LA, TX, MA, OH, AR, MT, MO, VA, NC, TN, CA, MN, WV, CT, CO, IA, MD, ID, NM, NH, UT, GA, WA, NV, OK, KS, AK, NE, HI.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 20, 2024
Recall initiation date
Feb 16, 2024
Center classification date
Mar 13, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.