Neodent GM X 6mm Engaging Titanium Base, REF: 9007162 — FDA record Z-1285-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 21, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Neodent GM X 6mm Engaging Titanium Base, REF: 9007162
Reason
Engaging titanium base contains a dimensional condition that allows vertical movement of the titanium base after the prosthetic screw is torqued to the recommended value, which if undetected and installed on the implant, may lead to abutment/screw fracture, which could result in the abutment/screw/crown being ingested by the patient.
Classification
Class II
Recalling firm
Preat Corp
Quantity
26
Agency-reported status
Ongoing
Source record identifier
Z-1285-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 00842092161326, Lots: 211891 and 211912

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of TX, OR, UT, NY.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 5, 2023
Recall initiation date
Feb 21, 2023
Center classification date
Mar 24, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.