ADVIA Centaur Erythropoietin (EPO) Assay (100 Test) — FDA record Z-1281-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 24, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ADVIA Centaur Erythropoietin (EPO) Assay (100 Test)
Reason
Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
264 units (6 US, 258 OUS)
Agency-reported status
Ongoing
Source record identifier
Z-1281-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySiemens Material Number: 10995096; UDI/DI: 00630414010380; Kit Lots: 53984040, 55443042.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AL, CA, IA, IL, MI, MN, NC, NE, NV, NY, OH and TX; The countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, P.R. China, Pakistan, Poland, Portugal, Republic of South Korea, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, U.A.E., United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 20, 2024
Recall initiation date
Jan 24, 2024
Center classification date
Mar 11, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.