Recall archive / FDA device enforcement
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 099... — FDA record Z-1277-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Feb 28, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- Reason
- There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.
- Classification
- Class II
- Recalling firm
- Datascope Corp.
- Quantity
- 391 total IABPs
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1277-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Model Number: 0998-00-0800-75, UDI-DI: 10607567112312; Model Number: 0998-00-0800-83, UDI-DI: 10607567108407; Model Number: 0998-00-0800-85, UDI-DI: 10607567113449; Serial Numbers: CA237026J3 CH231276C6 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Distribution list not yet provided.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Mar 29, 2023
- Recall initiation date
- Feb 28, 2023
- Center classification date
- Mar 22, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.