Colonovideoscope Model CF-H180AL — FDA record Z-1274-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 22, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Colonovideoscope Model CF-H180AL
Reason
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding
Classification
Class II
Recalling firm
Olympus Corporation of the Americas
Quantity
1 unit
Agency-reported status
Terminated
Source record identifier
Z-1274-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04953170339615 Serial number: 2501131

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of KY.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 29, 2022
Recall initiation date
Apr 22, 2022
Center classification date
Jun 21, 2022
Termination date
Mar 13, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.