Recall archive / FDA device enforcement
FreeStyle Libre Reader, REF: 71525-01, 71701-01, 71739-01 a component of the FreeStyle Libre Flash Glucose Monitoring... — FDA record Z-1273-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Feb 13, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- FreeStyle Libre Reader, REF: 71525-01, 71701-01, 71739-01 a component of the FreeStyle Libre Flash Glucose Monitoring System
- Reason
- Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.
- Classification
- Class I
- Recalling firm
- Abbott Diabetes Care, Inc.
- Quantity
- 336,085
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1273-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | REF/UDI-DI/Lot: 71525-01/357599000219/02Y702D, 01Y699D, 02Y699D, 01Y698D, 02Y697D, 04Y690D, 03Y690D, 01Y683D, 02Y683D, 01Y674D, 02Y672D, 01Y669D, 02Y666D, 01Y659D, 02Y658D, 02Y656D, 04Y655D, 01Y655D, 01Y652D, 02Y653D, 03Y653D, 03Y651D, 01Y641D, 02Y638D, 01Y637D, 02Y635D, 01Y629D, 03Y628D, 01Y628D, 02Y628D, 02Y621D, 02Y622D, 02Y620D, 02Y608D, 04Y608D, 01Y608D, 01Y603D, 03Y599D, 04Y600D, 02Y596D, 06Y596D, 03Y596D, 07Y547D, 07Y546D, 08Y546D, 05Y540D, 06Y538D, 05Y537D, 03Y536D, 04Y523D, 02Y522D, 02Y519D, 02Y512D, 01Y512D, 01Y511D, 02Y510D, 01Y505D, 02Y504D, 04Y355D, 02Y355D, 01Y352D, 01Y349D, 03Y348D, 01Y348D, 02Y348D, 01Y339D, 01Y335D, 02Y334D, 02Y335D, 03Y314D, 01Y313D, 02Y312D, 01Y312D, 02Y311D, 01Y311D, 02Y310D, 04Y310D, 02Y307D, 03Y307D, 01Y304D, 02Y303D, 02Y304D; 71701-01/357599000226/01Y702D, 01Y694D, 03Y693D, 02Y686D, 03Y683D, 02Y574D, 01Y563D, 02Y563D, 01Y564D, 03Y562D, 01Y552D, 02Y552D, 06Y550D, 05Y539D, 05Y538D, 06Y537D, 01Y536D, 01Y532D, 02Y531D, 03Y522D, 02Y518D, 02Y511D, 01Y508D, 02Y505D, 04Y356D, 01Y356D, 02Y356D, 01Y321D, 01Y320D, 02Y320D, 01Y317D, 02Y317D; 71739-01/357599000240/03Y754D, 03Y734D, 01Y705D, 01Y671D |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
U.S. Nationwide.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Apr 12, 2023
- Recall initiation date
- Feb 13, 2023
- Center classification date
- Apr 6, 2023