Alcon Clareon IOL with AutonoMe Delivery System — FDA record Z-1269-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 11, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Alcon Clareon IOL with AutonoMe Delivery System
Reason
Due to device (preloaded intraocular lens) mislabeling, associated with partial mixing of product during overlapping manufacturing runs, and attributed to improper manufacturing line clearance.
Classification
Class II
Recalling firm
Alcon Laboratories Ireland, Ltd
Quantity
139 lenses (Lot 25238769) and 164 lenses (Lot 25238775)
Agency-reported status
Terminated
Source record identifier
Z-1269-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalog/Model Number: CNA0T0.175 UDI Code: (01) 00380652393669 (17) 240720 (21) 25238769000 (11) 210722 (240) CNA0T0175 Lot number: 25238769 Catalog/Model Number: CNA0T0.240 UDI Code: (01) 00380652393799 (17) 240720 (21) 25238775000 (11) 210722 (240) CNA0T0240 Lot Number: 25238775

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S.: AL, CA, FL, GA, IA, IL, KY, LA, MI, MO, NY, OH,OK, TX, VA, and WI O.U.S.: Czech Republic, France, Germany, India, Slovakia, Spain, and Switzerland

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 29, 2022
Recall initiation date
May 11, 2022
Center classification date
Jun 17, 2022
Termination date
Feb 5, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.