CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU a... — FDA record Z-1265-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 10, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part numbers have been added as affected devices: (1) Part #2082294-001, FRU MP200 POWER SUPPLY ASSEMBLY; (2) Part #2082293-002-R, FRU MP200 CPU ASSEMBLY; (3) Part #2082293-002, FRU MP200 CPU ASSEMBLY; (4) Part #2071523-001, MP200 22 IN DISPLAY WITH TOUCH; (5) Part #2092930-001, DSP LCD 22 IN MED W/TCH W/DKST N AMR; (6) Part #2092930-002, DSP LCD 22 IN MED W/TCH W/DKST BRIT P/C; (7) Part #2092930-003, DSP LCD 22IN MED W/TCH W/DKST EURO P/C; (8) Part #2092930-004, DSP LCD 22IN MED W/TCH W/DKST ITA P/C; (9), Part #2092930-005, DSP LCD 22IN MED W/TCH W/DKST INDIAN P/C; (10) Part #2092930-006, DSP LCD 22IN MED W/TCH W/DKST ISREAL P/C; (11) Part #2092930-007, DSP LCD 22IN MED W/TCH W/DKST SWISS P/C; (12) Part #2092930-008, DSP LCD 22IN MED W/TCH W/DKST AUSTRL P/C; (13) Part #2092930-009, DSP LCD 22IN MED W/TCH W/DKST JPN; (14) Part #2092930-010, DSP LCD 22IN MED W/TCH W/DKST BRZL P/C; (15) Part #2092930-011, DSP LCD 22IN MED W/TCH W DKST ARGNT P C; (16) Part #2092930-012, DSP LCD 22IN MED W TCH W DKST DENMARK; (17) Part #2092930-013, DSP LCD22IN MED W TCH W DKST S AFRICA P C; and (18) Part #2092930-014, DSP LCD22IN MED W TCH W DKST CHINA P C.
Reason
The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.
Classification
Class II
Recalling firm
GE Healthcare, LLC
Quantity
30,993 Central Stations and 2,180 FRUs and displays
Agency-reported status
Ongoing
Source record identifier
Z-1265-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll CARESCAPE Central Station V2 units: (1) CSCS V2 MA1700 Integrated, Product Code SKN, Model #2082278-001, GTIN 00840682109666; (2) CSCS V2 MAS700 Desktop, Product Code SNF, Model #2082279-001, GTIN 00840682109604. NOTE: The Product Code is the first 3 letters of the serial number.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide. There was also military/government distribution and the countries of Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Kazakhstan, Kenya, Korea, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Macedonia, Madagascar, Malaysia, Maldives, Malta, Mauritius, Mexico, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, and Yemen.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 22, 2022
Recall initiation date
May 10, 2022
Center classification date
Jun 13, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records