Patient Side Cart, REF: 380620, with instrument arms, used as part of the da Vinci X Surgical System, REF: IS4200; Pa... — FDA record Z-1262-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jan 24, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Patient Side Cart, REF: 380620, with instrument arms, used as part of the da Vinci X Surgical System, REF: IS4200; Patient Side Cart, REF: 380652, with instrument arms, used as part of the da Vinci Xi Surgical System, REF: IS4000
- Reason
- Preventative maintenance data was used to identify instrument Arms, part of surgical systems, that allow surgeons to perform surgical tasks, that may be at risk of insertion ballscrew failure, allowing for uncontrolled carriage movement along the insertion axis, which may cause tissue injury.
- Classification
- Class II
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 9
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1262-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | System REF: Serial Numbers/Cart REF/UDI-DI/Arm Serial Numbers: IS4200: SL0379, SL0069/ 380620/ 00886874115404/ 668080, 575093, IS4000: SK1402, SK0991, SK2332, SK0744, SK0110, SK0753/380652 / 00886874110720/ 372054, 663693, 557322, 547953, 425439, 409962 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution in the states of FL, GA, WA and the countries of United Kingdom, Germany, France
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Mar 22, 2023
- Recall initiation date
- Jan 24, 2023
- Center classification date
- Mar 16, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.