In Vitro Diagnostic (IVD) Panel BioFire Respiratory Panel 2.1 (RP2.1) - REF 423742 — FDA record Z-1252-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 1, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
In Vitro Diagnostic (IVD) Panel BioFire Respiratory Panel 2.1 (RP2.1) - REF 423742
Reason
Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.
Classification
Class II
Recalling firm
BioFire Diagnostics, LLC
Quantity
8 kits
Agency-reported status
Ongoing
Source record identifier
Z-1252-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPart Number: 423742 UDI: 00815381020482 Pouch Lot Number: 1551622

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S. Nationwide distribution in the states of FL and GA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 22, 2023
Recall initiation date
Feb 1, 2023
Center classification date
Mar 15, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.