human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use. — FDA record Z-1251-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 18, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.
Reason
The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.
Classification
Class II
Recalling firm
HUMAN MED AG
Quantity
775 devices
Agency-reported status
Ongoing
Source record identifier
Z-1251-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot number 24-20198, UDI (01)04260170881014(11)240611(17)270611(10)24-20198; and Lot number 23-19612, UDI (01)04260170881014(11)240109(17)270109(10)23-19612.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of FL.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 12, 2025
Recall initiation date
Oct 18, 2024
Center classification date
Feb 28, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.