NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Model No. 300900 — FDA record Z-1251-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 13, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Model No. 300900
Reason
There is a potential for reduced analytical sensitivity for samples in the 1x - 3x LOD range for Flu A, RSV, and SARS-CoV-2 targets. This may lead to false negative results, allowing ongoing transmission to others and delaying appropriate treatment.
Classification
Class II
Recalling firm
NeuMoDx Molecular Inc
Quantity
459 kits US; 1626 kits OUS
Agency-reported status
Ongoing
Source record identifier
Z-1251-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI (GTIN): 10814278020830 Cartridge lot numbers 111915 - 117508

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution to FL, NY, MD, PA, VA, MI, TN, NM. Foreign distribution to AT, AU, BE, CH, DE, FI, FR, GB, GR, HK, ID, IT, LT, NL, SA, SE, and SI.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 15, 2022
Recall initiation date
May 13, 2022
Center classification date
Jun 8, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records