Plato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090 — FDA record Z-1249-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 1, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Plato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090
Reason
Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.
Classification
Class II
Recalling firm
Scientia Vascular, Inc.
Quantity
353
Agency-reported status
Ongoing
Source record identifier
Z-1249-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll lots, REF/UDI-DI/Lot(Expiration): PL17-160-000/00818075010455/ 900164(8/31/2022), 900165(8/31/2022), 900172(8/31/2022), 900177(9/30/2022), 900180(9/30/2022), 900186(10/31/2022), 900187(10/31/2022), 900188(10/31/2022), 900193(10/31/2022), 900196(12/31/2022), 900199(12/31/2022), 900202(12/31/2022), 900221(2/28/2023), 900222(2/28/2023), 900225(2/28/2023), 900226(3/31/2023), 20441(11/30/2023), 20476(11/30/2023), 20650(12/31/2023); PL17-160-045/00818075010462/900167(8/31/2022), 900168(8/31/2022), 900169(8/31/2022), 900174(8/31/2022), 900175(8/31/2022), 900179(9/30/2022), 900181(9/30/2022), 900183(10/31/2022), 900185(10/31/2022), 900205(1/31/2023), 900206(1/31/2023), 900210(1/31/2023), 900211(1/31/2023), 900215(1/31/2023), 900220(1/31/2023), 900219(1/31/2023), 900217(2/28/2023), 900223(2/28/2023), 900224(2/28/2023); PL17-160-090/00818075010479/900170(8/31/2022), 900171(8/31/2022), 900178(9/30/2022), 900182(9/30/2022), 900184(10/31/2022), 900192(11/30/2022), 900195(11/30/2022), 900209(1/31/2023), 900228(3/31/2023)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of IL, AL, FL, NC, WI, GA, SC, IN, KS, CA, OH, NY, NH, CO, MA, NV, UT, TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 22, 2023
Recall initiation date
Feb 1, 2023
Center classification date
Mar 13, 2023