Focalyx Fusion — FDA record Z-1243-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 23, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Focalyx Fusion
- Reason
- Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.
- Classification
- Class II
- Recalling firm
- Focalyx Technologies, LLC.
- Quantity
- 18
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1243-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution in the states of NY, MA, FL and the countries of ES, VE, DO.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Feb 11, 2026
- Recall initiation date
- Dec 23, 2025
- Center classification date
- Feb 3, 2026