Optima MR360, NMRI system — FDA record Z-1241-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 31, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Optima MR360, NMRI system
Reason
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Classification
Class II
Recalling firm
GE Healthcare (China) Co., Ltd.
Quantity
33
Agency-reported status
Ongoing
Source record identifier
Z-1241-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityA. UDI-DI 00840682113762 System ID 0910274038 B. No UDI-DI System IDs 082427040134 082427040185 082427060043 082427070088 082427070093 082427070099 082427100174 082427100178 082427100197 082427120140 082427140095 082427220046 082427270041 082427310088 082427310092 083027288790310 0856270061 AE1373MR01 DZ1339MR03 FI1432MR01 H4678MR04 MRR9282 MRR9877 PL2059MR01 PL2433MR01 RU3338MR03 SA1056MR03 SA2191MR04 SA2253MR01 SA2352MR01 TN1008MR01 TN1241MR01 34352MRS01 (Update 12/29/25 - Additional System IDs 83027809699615, LV4026MR01) (Update 1/28/26 - Additional System ID 82427190104)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic US distribution nationwide. International distribution worldwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 5, 2025
Recall initiation date
Jan 31, 2025
Center classification date
Feb 27, 2025