DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software. — FDA record Z-1233-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 10, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.
Reason
Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.
Classification
Class II
Recalling firm
BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD.
Quantity
56
Agency-reported status
Ongoing
Source record identifier
Z-1233-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySoftware Versions: V1.4, V1.4.1, V1.4.2, V1.4.3 REF/UDI-DI/Serial Numbers: C63519/14987666545058/2024090170; C63520/14987666545065/2024070116, 2023070009, 2024010038, 2024010035, 2024050072, 2024040056, 2024030049, 2024020046, 2024070111, 2024030051, 2024030052, 2024010033, 2023090015, 2024060103, 2024060104, 2024060101, 2024080145, 2024020045, 2024020042, 2024020043, 2024020047, 2024060096, 2024040059, 2023100019, 2024030053, 2023070010, 2024070129, 2024070130, 2024020041, 2024090183, 2024010036, 2024010037, 2024070132, 2024070131, 2023100020, 2024050064, 2024050070, 2023090016, 2023110022, 2024070110, 2024060075, 2024060106, 2023110023, 2024010024, 2024010025, 2024050066, 2024010029, 2024060107, 2023080012, 2024040060, 2024050074, 2024040062, 2024010031, 2024030055, 2024080144

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: MA, CA, SC, LA, OK, GA, AL, TX, OR, ID, CO, MN, NY, MS, PA, WA, IL, MI, PR. OUS: Lebanon, Poland, Ghana, Spain, South Africa, Slovakia, Ghana, Croatia, Seychelles, Australia, Republic of Korea

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 5, 2025
Recall initiation date
Jan 10, 2025
Center classification date
Feb 25, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records