Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-0... — FDA record Z-1231-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Feb 7, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D998-00-0800-53, D998-00-0800-55, D998-00-0800-65, D998-00-0800-83, D998-UC-0800-31, D998-UC-0800-33, D998-UC-0800-52, D998-UC-0800-53, D998-UC-0800-55.
- Reason
- Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
- Classification
- Class II
- Recalling firm
- Datascope Corp.
- Quantity
- 11,792 units (5.408 US, 6,384 OUS)
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1231-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Model Number (UDI-DI): D998-00-0800-31 (10607567109053¿), D998-00-0800-32 (10607567111117), D998-00-0800-33 (10607567109008), D998-00-0800-35 (10607567109107), D998-00-0800-36 (10607567114187), D998-00-0800-45 (10607567108421), D998-00-0800-52 (10607567108438), D998-00-0800-53 (10607567108391), D998-00-0800-55 (10607567108414), D998-00-0800-65 (10607567113432), D998-00-0800-83 (10607567108407), D998-UC-0800-31 (10607567109053), D998-UC-0800-33 (10607567109008), D998-UC-0800-52 (10607567108438), D998-UC-0800-53 (10607567108391), D998-UC-0800-55 (10607567108414). All serial numbers |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Nationwide US distribution. Worldwide international distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Mar 15, 2023
- Recall initiation date
- Feb 7, 2023
- Center classification date
- Mar 6, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.