Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, ... — FDA record Z-1229-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 7, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, and 0998-00-0800-55
Reason
There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.
Classification
Class II
Recalling firm
Datascope Corp.
Quantity
8909 units
Agency-reported status
Ongoing
Source record identifier
Z-1229-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll Serial Numbers. Model Number: 0998-00-0800-31, UDI-DI:10607567109053; Model Number: 0998-00-0800-32, UDI-DI: 10607567111117; Model Number: 0998-00-0800-33, UDI-DI: 10607567109008; Model Number: 0998-00-0800-34, UDI-DI: 10607567111940; Model Number: 0998-00-0800-35, UDI-DI: 10607567109107; Model Number: 0998-00-0800-45, UDI-DI: 10607567108421; Model Number: 0998-00-0800-52, UDI-DI: 10607567108438; Model Number: 0998-00-0800-53, UDI-DI: 10607567108391; Model Number: 0998-00-0800-55, UDI-DI: 10607567108414; Model Number: 0998-00-0800-65, UDI-DI: 10607567113432

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide. Global Distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 15, 2023
Recall initiation date
Feb 7, 2023
Center classification date
Mar 6, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.