VITROS XT 7600 Integrated System, Catalog No. 6844461 — FDA record Z-1228-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 26, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
VITROS XT 7600 Integrated System, Catalog No. 6844461
Reason
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Classification
Class II
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
44 US; 35 OUS
Agency-reported status
Ongoing
Source record identifier
Z-1228-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityVITROS System Software Version 3.8.1 UDI-DI 10758750031610 Serial Numbers US: J76000199 J76001383 J76000207 J76001385 J76000353 J76001392 J76000456 J76001404 J76000459 J76001406 J76000460 J76001516 J76000468 J76001518 J76000495 J76001519 J76000502 J76001520 J76000561 J76001581 J76001207 J76001583 J76001230 J76001611 J76001231 J76001620 J76001232 J76001644 J76001235 J76001647 J76001239 J76001660 J76001260 J76001668 J76001263 J76001671 J76001270 J76001674 J76001274 J76001692 J76001331 J76001700 J76001351 J76001748 OUS: J76000118 J76001070 J76000253 J76001074 J76000265 J76001095 J76000274 J76001121 J76000281 J76001130 J76000370 J76001152 J76000447 J76001174 J76000475 J76001328 J76000620 J76001377 J76000855 J76001378 J76000856 J76001396 J76000898 J76001399 J76000901 J76001407 J76000973 J76001408 J76000976 J76001484 J76001001 J76001673 J76001027 J76001761 J76001035

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 13, 2024
Recall initiation date
Jan 26, 2024
Center classification date
Mar 1, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.