BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated plasma. Pr... — FDA record Z-1227-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 1, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated plasma. Protein C is a glycoprotein, vitamin K dependent, which inhibits coagulation. Protein C in plasma is measured following a specific activation with Protac (Agkistrodom Contortrix), an enzyme extracted from snake venom. Activated protein C hydrolysis the chromogenic substrate (SaPC-21) which release para-nitroaniline (pNA). The amount of pNA released (measured by absorbance at 405 nm) is directly proportional to the concentration of Protein C in the specimen.
Reason
New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to the CLP regulation ((EC) No 1272/2008) is suspected of damaging fertility
Classification
Class II
Recalling firm
Aniara Diagnostica LLC
Quantity
2 units
Agency-reported status
Ongoing
Source record identifier
Z-1227-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 366353700063BD Lot FA083327 (And all future lots)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of NC.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 15, 2023
Recall initiation date
Feb 1, 2023
Center classification date
Mar 6, 2023