VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished) — FDA record Z-1224-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 26, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished)
Reason
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Classification
Class II
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
1 US; 36 OUS
Agency-reported status
Ongoing
Source record identifier
Z-1224-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityVITROS System Software Version 3.8.1 Regular UDI-DI 10758750002979 Refurbished UD-DI 10758750007103 Serial Numbers US: J36001167 OUS: J36000427 J36001117 J36001558 J36000580 J36001154 J36001570 J36000626 J36001229 J36001573 J36000688 J36001427 J36001575 J36000877 J36001451 J36001579 J36000920 J36001488 J36001582 J36000929 J36001489 J36001583 J36000942 J36001495 J36001595 J36001012 J36001519 J36001602 J36001051 J36001528 J36001608 J36001108 J36001530 J36001632 J36001111 J36001535 J36001640

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 13, 2024
Recall initiation date
Jan 26, 2024
Center classification date
Mar 1, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.