NIM Vital Console 4.0 (P/N: NIM4CM01) NIM with Software version 1.3.2 — FDA record Z-1223-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 3, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
NIM Vital Console 4.0 (P/N: NIM4CM01) NIM with Software version 1.3.2
Reason
Software anomaly was identified.
Classification
Class II
Recalling firm
Medtronic Xomed, Inc.
Quantity
18 units
Agency-reported status
Ongoing
Source record identifier
Z-1223-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel Number: NIM4CM01; UDI/DI: 00763000395896, lot numbers: C2026207, C2026233, C2026320, C2026335, C2026354, C2026376, C2110046, C2110089, C2110131, C2110196, C2110198, C2110208, C2110217, C2110223 and 00763000002978, lot numbers: C1915048, C1915078, C1915099, C1915188, C2026207, C2026233, C2026320, C2026335, C2026354, C2026376, C2110046, C2110089, C2110131, C2110196, C2110198, C2110208, C2110217, C2110223.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of TX, CO, NJ and the country of France.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 8, 2022
Recall initiation date
May 3, 2022
Center classification date
Jun 2, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records