Colonoscope, Model Number PCF-H190DL. — FDA record Z-1210-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jan 11, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Colonoscope, Model Number PCF-H190DL.
- Reason
- Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
- Classification
- Class II
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 47
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1210-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Model Number: PCF-H190DL; UDI DI: 04953170363672; Serial Number: 2046148, 2046377, 2500549, 2501714, 2602501, 2602549, 2602675, 2602884, 2603086, 2605185, 2605189, 2605279, 2605338, 2605343, 2615621, 2716237, 2716306, 2716379, 2727123, 2727529, 2737897, 2738603, 2839269, 2839284, 2839580, 2840038, 2840130, 2840639, 2840669, 2840719, 2840798, 2840928, 2842302, 2942674, 2942814, 2943131, 2943193, 2943240, 2943296, 2943451, 2943708, 2943820, 2944426, 2944690, 2944765, 2944939, 2945260; |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Mar 6, 2024
- Recall initiation date
- Jan 11, 2024
- Center classification date
- Feb 29, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.