Recall archive / FDA device enforcement
Colonoscope, Model Number CF-H180AL. — FDA record Z-1205-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jan 11, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Colonoscope, Model Number CF-H180AL.
- Reason
- Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
- Classification
- Class II
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 36
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1205-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Model Number: CF-H180AL; UDI DI: 4953170339608; Serial Number: 2001426, 2001524, 2001614, 2002139, 2002212, 2002285, 2103817, 2104137, 2104581, 2105255, 2105976, 2107112, 2207880, 2207929, 2207940, 2309317, 2601078, 2601096, 2601255, 2601368, 2601700, 2702380, 2702460, 2702478, 2702805, 2703697, 2704366, 2704490, 2806996, 2807975, 2900004, 2900185, 2900249, 2908770, 2909110, 2909884; |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Mar 6, 2024
- Recall initiation date
- Jan 11, 2024
- Center classification date
- Feb 29, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.