Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OE-001 — FDA record Z-1204-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 27, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OE-001
Reason
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be triggered, the initial alarm/alert will not be delivered until a subsequent alarm/alert is triggered. At that time, the initial alarm/alert is delivered, but not the second one. Alarm/alerts will continue to be delayed and be one alarm/alert behind, causing delayed alarm/alerts until the receiver is reset. This condition could result in the missed detection of a hyperglycemic or hypoglycemic event. Second, a single EGV reading may be delayed for 5 minutes after the initial .net operating system error. If the user receives a single EGV reading delayed by 5 minutes, the delay is not expected to cause user harm.
Classification
Class II
Recalling firm
Dexcom, Inc.
Quantity
N/A
Agency-reported status
Ongoing
Source record identifier
Z-1204-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll Serial Numbers/UDI: 00386270001900

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of OH, TX, PA, MN, SC, NH, FL, CA, NJ, PR, GA and the countries of Austria, Belgium, Canada, Switzerland, Germany, Spain, France, Ireland, Japan, Netherlands, Slovenia, United Kingdom, Australia, Czech Republic, Denmark, Finland, France, Hong Kong, Israel, Italy, Korea, Kuwait, New Zealand, Norway, Oman, Poland, Qatar, Saudi Arabia, Singapore, Slovakia, South Africa, Sweden, Turkey, United Arab Emirates.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 26, 2025
Recall initiation date
Jan 27, 2025
Center classification date
Feb 20, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records