Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards, with IP addresses from x.x.x.224 to ... — FDA record Z-1192-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 24, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards, with IP addresses from x.x.x.224 to x.x.x.255
Reason
Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed
Classification
Class II
Recalling firm
CareFusion 303, Inc.
Quantity
83,185
Agency-reported status
Ongoing
Source record identifier
Z-1192-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll PC Units, UDI-DI: 10885403801549, 10885403801532, with software versions 12.1.0, 12.1.2, 12.1.3, utilizing affected network cards and IP addresses

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: FL, CA, OR, IL, WI, NJ, AZ, NV, CO, WY, NE, TX, MI, VA, OH, NC, LA, NM, AL, TN, AR, PA, CT, IN, KS, GA, SC, MD, UT, MA, DC, OK, IA, NY, WA, SD, ND, MN, KY, MO, WV, MS, DE, ME, MT, ID, NH, HI, AK, VT OUS: Canada, Qatar, Saudi Arabia,, Kuwait, United Arab Emirates, Israel, South Africa, Zimbabwe, Botswana, Australia, New Zealand

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 8, 2023
Recall initiation date
Jan 24, 2023
Center classification date
Mar 1, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.