BALLARD ACCESS Closed Suction System for Neonates/Pediatrics, Y-Manifold. — FDA record Z-1189-2023
Reported terminated
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Feb 3, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- BALLARD ACCESS Closed Suction System for Neonates/Pediatrics, Y-Manifold.
- Reason
- Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation.
- Classification
- Class I
- Recalling firm
- Avanos Medical, Inc.
- Quantity
- 780 units
- Agency-reported status
- Terminated
- Source record identifier
- Z-1189-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | All lots are affected. Stock Code (Lot Number): B30511 (20098360), B30611 (20098361), B30711 (20097200), B30811 (20098362), B31011 (20098363). |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the state of Oklahoma.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Mar 22, 2023
- Recall initiation date
- Feb 3, 2023
- Center classification date
- Mar 13, 2023
- Termination date
- Sep 5, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.