Recall archive / FDA device enforcement
Rebstock Holding rod for fixation -For holding and fixation of flexible Leyla Arms and spatula Art. No. 06-06-285 and... — FDA record Z-1184-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 21, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Rebstock Holding rod for fixation -For holding and fixation of flexible Leyla Arms and spatula Art. No. 06-06-285 and 06-06-285MOD. surgical instrument.
- Reason
- Holding arm could move (twist) uncontrollably in the front area. This could lead to a serious deterioration of patient's safety if the movement is transferred to the spatula, which is attached to the holding arm.
- Classification
- Class II
- Recalling firm
- Medagent Inc
- Quantity
- 84 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1184-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Batches: 6221672, 6222079, 6230282, 6230985, 6230986, 6231808, 6222559, 6231989, 6221673 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
TX Foreign: Algeria China Dominican Republic Germany India Japan Philippines UK
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Mar 6, 2024
- Recall initiation date
- Dec 21, 2023
- Center classification date
- Feb 23, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.