Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive s... — FDA record Z-1177-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 6, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228
Reason
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
Classification
Class II
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Quantity
27 units
Agency-reported status
Ongoing
Source record identifier
Z-1177-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityOUS (1) Model: 722227 SN: UDI: 75 (01)00884838099227(21)75 122 (01)00884838099227(21)122 131 (01)00884838099227(21)131 132 (01)00884838099227(21)132 165 (01)00884838099227(21)165 116 (01)00884838099227(21)116 136 (01)00884838099227(21)136.' (2) Model: 722228 Serial Number UDI 168 (01)00884838099234(21)168 197 (01)00884838099234(21)197 154 (01)00884838099234(21)154 193 (01)00884838099234(21)193 165 (01)00884838099234(21)165 163 (01)00884838099234(21)163 139 (01)00884838099234(21)139 202 (01)00884838099234(21)202 141 (01)00884838099234(21)141 200 (01)00884838099234(21)200 174 (01)00884838099234(21)174 178 (01)00884838099234(21)178 147 (01)00884838099234(21)147 208 (01)00884838099234(21)208 171 (01)00884838099234(21)171 167 (01)00884838099234(21)167 194 (01)00884838099234(21)194 176 (01)00884838099234(21)176 228 (01)00884838086777(10)2.0.0.10015 161 (01)00884838099234(21)161

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Macao, Macedonia, Malaysia, Morocco, Netherlands, Norway, Oman, Palestine, State of Panama, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, Ukraine, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 28, 2024
Recall initiation date
Feb 6, 2024
Center classification date
Feb 22, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.