3M Specialty Blade Assembly, REF 9660,70-2011-8462-2, 7100213114, used with used with 3M Surgical Clipper Catalog Num... — FDA record Z-1169-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 12, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
3M Specialty Blade Assembly, REF 9660,70-2011-8462-2, 7100213114, used with used with 3M Surgical Clipper Catalog Number 9661L
Reason
3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41¿C (105.8¿F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60¿C (140¿F) and the Model 9660 blade maximum temperature may reach 77¿C (170.6¿F)."
Classification
Class II
Recalling firm
3M Company - Health Care Business
Quantity
52,833,750 blades
Agency-reported status
Ongoing
Source record identifier
Z-1169-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 50707387789003 (shipper), 30707387789009 (tray), all lot codes

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution.

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Feb 28, 2024
Recall initiation date
Jan 12, 2024
Center classification date
Feb 21, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.