ESG PK CUTTING FORCEPS, 5MM, 33CM — FDA record Z-1167-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 15, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ESG PK CUTTING FORCEPS, 5MM, 33CM
Reason
The jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure.
Classification
Class II
Recalling firm
Olympus Corporation of the Americas
Quantity
11,123 units
Agency-reported status
Ongoing
Source record identifier
Z-1167-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel: PK-CF0533; UDI DI: 00821925035867; Lot Numbers: FR258486, FR261738, FR259544, FR263183, FR276278, FR279757, FR279625, FR279658, FR286369, FR287112, FR374622, FR207935, FR212413, FR212735, FR215061, FR216817, FR220525, FR222456, FR227254, FR227774, FR227781, FR232882, FR234845, FR247614, FR240430, FR246627, FR247039, FR248008, FR255679, FR248312, FR263636, FR263244, FR269280, FR269267, FR269269, FR287114, FR303799, FR297022, FR301062, PWO-308561, PWO-308562, PWO-308563, PWO-308560, PWO-308567, PWO-308566, PWO-308565, FR303864, FR313639, FR316869, FR319703, FR316872, FR327890, PWO-308972, FR335984, FR335990, FR337947, FR373912, FR401646, FR379790, FR347006;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Mexico, P.R. China, Singapore, South Korea.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 28, 2024
Recall initiation date
Dec 15, 2023
Center classification date
Feb 20, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records