Luminos dRF with software VD10- A diagnostic imaging system for radiographic and fluoroscopic studies Model Number: 1... — FDA record Z-1166-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 21, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Luminos dRF with software VD10- A diagnostic imaging system for radiographic and fluoroscopic studies Model Number: 10094200
Reason
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
1 unit; Expanded Recall: 25 units
Agency-reported status
Ongoing
Source record identifier
Z-1166-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: N/A S/N: 4650 Expanded Recall 2/24/23: 4043 4051 4143 4156 4199 4201 4238 4252 4291 4323 4332 4342 4354 4372 4401 4430 4508 4600 4606 4637 4641 4683 4722 4724 4734

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 1, 2023
Recall initiation date
Dec 21, 2022
Center classification date
Feb 23, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.