The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time.... — FDA record Z-1155-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 15, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.
Reason
VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.
Classification
Class II
Recalling firm
Canon Medical System, USA, INC.
Quantity
2
Agency-reported status
Ongoing
Source record identifier
Z-1155-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04987670100208. Serial Number: BGE2292016, BGE2362017

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: CA, LA

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 28, 2024
Recall initiation date
Nov 15, 2023
Center classification date
Feb 16, 2024