Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Nu... — FDA record Z-1146-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 7, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Reason
An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.
Classification
Class I
Recalling firm
Datascope Corp.
Quantity
4062 Total
Agency-reported status
Ongoing
Source record identifier
Z-1146-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll serial numbers. Model Number 0998-00-0800-75, UDI-DI 10607567112312; Model Number 0998-00-0800-83, UDI-DI 10607567108407; Model Number 0998-00-0800-85, UDI-DI 10607567113449

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, BRUNEI , BULGARIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GHANA, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, TURKMENISTAN, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM, AND YEMEN.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 15, 2023
Recall initiation date
Feb 7, 2023
Center classification date
Mar 9, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records