HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-... — FDA record Z-1141-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 14, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-10; SAM 360P/360-STR-JA-10, 360-STR-DE-10, 360-STR-UK-10; SAM 450P/450-UNIT-US-08, ; SAM 500P/500-BAS-UK-10
Reason
Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.
Classification
Class II
Recalling firm
HeartSine Technologies Ltd
Quantity
12
Agency-reported status
Ongoing
Source record identifier
Z-1141-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityType/UDI-DI or GTIN/Serial Number: SAM 350P/05060167120671, 05060167127670, +M727SAM350P0/$$+7/17D00023014, 18D00020508, 19D00017655, 20D00006524, 21D91081653, 22D91152391; SAM 360P/05060167127687, 05060167127311,5060167122927/21E91049810, 23E90001379, 23E90010409; SAM 450P/05060167129773, +M727SAM450P0/$$+7, +M727SAM450P0/$$+7/17G00001893, 18G00001553; SAM 500P/05060167122453/22B91071233

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of TX, CA, IL, CO, FL, NC, MI and the countries of Netherlands, Japan, New Zealand.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 21, 2024
Recall initiation date
Feb 14, 2024
Center classification date
Feb 15, 2024