What the record says
- Product
- HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-10; SAM 360P/360-STR-JA-10, 360-STR-DE-10, 360-STR-UK-10; SAM 450P/450-UNIT-US-08, ; SAM 500P/500-BAS-UK-10
- Reason
- Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.
- Classification
- Class II
- Recalling firm
- HeartSine Technologies Ltd
- Quantity
- 12
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1141-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Type/UDI-DI or GTIN/Serial Number: SAM 350P/05060167120671, 05060167127670, +M727SAM350P0/$$+7/17D00023014, 18D00020508, 19D00017655, 20D00006524, 21D91081653, 22D91152391; SAM 360P/05060167127687, 05060167127311,5060167122927/21E91049810, 23E90001379, 23E90010409; SAM 450P/05060167129773, +M727SAM450P0/$$+7, +M727SAM450P0/$$+7/17G00001893, 18G00001553; SAM 500P/05060167122453/22B91071233 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution in the states of TX, CA, IL, CO, FL, NC, MI and the countries of Netherlands, Japan, New Zealand.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Feb 21, 2024
- Recall initiation date
- Feb 14, 2024
- Center classification date
- Feb 15, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.