Mazor X robotic guidance system REF: TPL0059 — FDA record Z-1122-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 10, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Mazor X robotic guidance system REF: TPL0059
Reason
Software errors that can result in incorrect surgical instrument positioning during spinal surgery.
Classification
Class II
Recalling firm
Mazor Robotics Ltd
Quantity
549 systems
Agency-reported status
Ongoing
Source record identifier
Z-1122-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySoftware versions: 5.0.1, 5.1.2, 5.1.3/ UDI: 07290109180465, 07290109184524, 07290109181158, 00763000635169, 07290109183213, 07290109184517, 00763000419134, 00763000431761, 07290109184098, 07290109184838, 07290115751376, 07290115751895

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, China, Croatia, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States, Vietnam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 28, 2026
Recall initiation date
Dec 10, 2025
Center classification date
Jan 16, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.