MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0... — FDA record Z-1121-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 11, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended purpose of the LVIS Device is to serve as a stent-assisted coiling (SAC) device providing support to the embolic coils used in the treatment of wide-necked aneurysms.
Reason
Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.
Classification
Class II
Recalling firm
MICROVENTION INC.
Quantity
11 units
Agency-reported status
Ongoing
Source record identifier
Z-1121-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot number: 0000456768

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: None OUS: China

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Feb 19, 2025
Recall initiation date
Dec 11, 2024
Center classification date
Feb 7, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.