VITROS Immunodiagnostic Products FSH Reagent Pack- IVD 1931922 — FDA record Z-1112-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 25, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
VITROS Immunodiagnostic Products FSH Reagent Pack- IVD 1931922
Reason
Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.
Classification
Class II
Recalling firm
Ortho-Clinical Diagnostics, Inc
Quantity
2502 total: 2154 US; 348 OUS
Agency-reported status
Ongoing
Source record identifier
Z-1112-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUntil further notice, future lots will also be affected. UDI: 10758750000302 Affected Lot Number/ Expiration Date: 3100 01-MAR-2022; 3111 25-MAR-2022; 3120 04-APR-2022; 3130 25-APR-2022; 3140 10-MAY-2022; 3145 10-MAY-2022; 3150 30-MAY-2022; 3160 09-JUN-2022; 3170 22-JUN-2022; 3180 02-AUG-2022; 3190 16-AUG-2022; 3200 06-SEP-2022; 3210 12-SEP-2022.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia Singapore Spain Sweden United Kingdom

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 25, 2022
Recall initiation date
Mar 25, 2022
Center classification date
May 18, 2022